Coupling matrix seems to be always loaded when using FIXED tool-detection mode #203

Closed
yafei-ou opened this issue Apr 22, 2024 · 0 comments

Comments

@yafei-ou
Copy link

yafei-ou commented Apr 22, 2024

After a meeting with Anton @adeguet1, we have found a bug in the software version 2.2.1.

When using FIXED tool-detection mode, the coupling matrix seems to be always loaded. This affects the engagement of the sterile adaptor and also the calibration. When engaging the sterile adaptor, it seems that the actuators for the disk will go out of the limits because of the matrix being present. When calibrating the potentiometers, as the coupling matrix is present, the values are wrong.

Current way of mitigation: avoid using FIXED mode for tool-detection, either by using AUTOMATIC or MANUAL.