Skip to content

Update TestComponent.txt #55

New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

Open
wants to merge 1 commit into
base: master
Choose a base branch
from
Open
Changes from all commits
Commits
File filter

Filter by extension

Filter by extension

Conversations
Failed to load comments.
Loading
Jump to
Jump to file
Failed to load files.
Loading
Diff view
Diff view
38 changes: 38 additions & 0 deletions examples/MVLinac/TestComponent.txt
Original file line number Diff line number Diff line change
Expand Up @@ -20,4 +20,42 @@ includeFile = Flattener18MV.txt #Comment out at least one of the flatteners
#includeFile = WaterBox.txt #Comment out at least one of the dose scoring regions
#includeFile = WaterPhantom.txt #Comment out at least one of the dose scoring regions

# Source selection (electron or x-ray) and number of histories in run
ic:So/ElectronSource/NumberOfHistoriesInRun = 100
#ic:So/XraySource/NumberOfHistoriesInRun = 100

# Source at nominal position of target
# Parent is the gantry with coordinates IEC_G (see Main.txt)
s:Ge/BeamPosition/Parent = "IEC_G"
dc:Ge/BeamPosition/Pos = -1.0 cm
d:Ge/BeamPosition/TransZ = Ge/SAD - Ge/BeamPosition/Pos cm

# Electron source
s:So/ElectronSource/Type = "Beam"
s:So/ElectronSource/Component = "BeamPosition"
s:So/ElectronSource/BeamParticle = "e-"
d:So/ElectronSource/BeamEnergy = 6.51 MeV #6 MV Oncor
u:So/ElectronSource/BeamEnergySpread = 4.2 #10% FWHM 6 MV Oncor
s:So/ElectronSource/BeamPositionDistribution = "Gaussian"
s:So/ElectronSource/BeamPositionCutoffShape = "Ellipse"
d:So/ElectronSource/BeamPositionCutoffX = 1.2 mm #2 standard deviations
d:So/ElectronSource/BeamPositionCutoffY = 1.2 mm
d:So/ElectronSource/BeamPositionSpreadX = 0.6 mm #1.5 mm FWHM
d:So/ElectronSource/BeamPositionSpreadY = 0.6 mm
s:So/ElectronSource/BeamAngularDistribution = "None"

# Monoenergetic x-ray source
s:So/XraySource/Type = "Beam"
s:So/XraySource/Component = "BeamPosition"
s:So/XraySource/BeamParticle = "gamma"
d:So/XraySource/BeamEnergy = 2 MeV
u:So/XraySource/BeamEnergySpread = 0.1
s:So/XraySource/BeamPositionDistribution = "Gaussian"
s:So/XraySource/BeamPositionCutoffShape = "Ellipse"
d:So/XraySource/BeamPositionCutoffX = 1.2 mm #2 standard deviations
d:So/XraySource/BeamPositionCutoffY = 1.2 mm
d:So/XraySource/BeamPositionSpreadX = 0.6 mm #1.5 mm FWHM
d:So/XraySource/BeamPositionSpreadY = 0.6 mm
s:So/XraySource/BeamAngularDistribution = "Flat"
d:So/XraySource/BeamAngularCutoffX = 14. deg #50 cm diameter collimator
d:So/XraySource/BeamAngularCutoffY = 14. deg